FDA issues Class I recall for Philips ultrasound

  • By

  • Teraya Smith

  • October 20, 2026

  • 2 min

Share

Clinical Scorecard: FDA issues Class I recall for Philips ultrasound

At a Glance

CategoryDetail
ConditionPhilips Lumify diagnostic ultrasound system functionality issues
Key MechanismsSoftware may unexpectedly require transducer re-registration, which requires Wi-Fi or cellular connectivity; unstable or disconnected USB connections may interrupt imaging.
Target PopulationPatients examined using affected Philips Lumify systems.
Care SettingPoint-of-care ultrasound, including ambulances, aircraft, and field settings where network access may be limited.

Key Highlights

  • The FDA classified the correction as a Class I recall on October 19, following an October 9 early alert.
  • As of September 29, 2026, 8 serious injuries and 4 deaths were associated with device functionality issues.
  • Unexpected transducer re-registration can prevent scanning until registration is completed using Wi-Fi or cellular connectivity.
  • USB connection problems may interrupt continuous imaging and delay care.
  • The recall involves updated operating instructions rather than device removal.

Guideline-Based Recommendations

Diagnosis

    Management

    • Philips instructed customers to use Lumify systems only where Wi-Fi or cellular connectivity is readily accessible, unless a backup ultrasound device is immediately available.
    • Philips recommended maintaining alternative imaging capabilities and distributing safety notices to device users.

    Monitoring & Follow-up

      Risks

      • Loss of ultrasound functionality may interfere with rapid diagnosis, treatment decisions, and triage of unstable patients.
      • Depending on clinical circumstances, resulting delays could cause serious injury or death.

      Patient & Prescribing Data

      Patients examined using affected Lumify systems; the article does not provide patient-level demographic or prescribing data.

      The article describes device-use precautions, not patient treatment regimens. Philips is developing a software update for unexpected re-registration and separately evaluating the USB connectivity problem.

      Clinical Best Practices

      • Use affected Lumify systems only where Wi-Fi or cellular connectivity is readily accessible, unless a backup ultrasound device is immediately available.
      • Maintain alternative imaging capabilities.
      • Distribute the safety notice to device users.