Putting a POC Troponin to the Test - Summary - MDSpire
Coming Soon: Introducing MDSpire News. Learn more
Conexiant’s news site is now MDSpire News. Learn more

Putting a POC Troponin to the Test

  • By

  • Julie Greenbaum

  • February 5, 2026

  • 3 min

Share

Objective:

To verify the analytical performance and concordance of the QuidelOrtho TriageTrue point-of-care high-sensitivity troponin I assay in patients assessed for myocardial infarction.

Approach:
    Key Findings:
    • Intra-assay precision CVs were 25% and 12.2% at mean concentrations of 2.4 ng/L and 8.5 ng/L, respectively, with overall performance metrics indicating adequacy.
    • Hemolysis caused minimal interference up to 10 g/L, exceeding the manufacturer's claim.
    • Correlation coefficients with other assays ranged from 0.89 to 0.99, except for Roche hs-TnT assay (0.79).
    • All 116 comparisons with secondary analyzers met acceptable limits.
    Interpretation:

    The hs-TnI assay on the TriageTrue demonstrates adequate performance for clinical use, supporting its integration into clinical pathways based on the findings.

    Limitations:
    • Logistical constraints in obtaining suitable whole blood EDTA samples limited targeted sampling at specific troponin I levels, impacting the study's findings.
    Conclusion:

    The assay's analytical performance supports its use in emergency care and low-resource settings, aligning with manufacturer claims and emphasizing its clinical significance.

    Sources:

Original Source(s)

Related Content