To evaluate the clinical feasibility and technical performance of augmented reality (AR) navigation for brain glioma resection in a preliminary patient series.
Key Findings:
Gross total resection of the contrast-enhancing tumor was achieved in all patients.
Mean AR registration time was 4.7 ± 0.7 minutes.
Surface-matching registration yielded a TRE of 2.1 ± 0.4 mm.