Objective:

To validate VisionMetric Suite, an AI-based automated periocular measurement tool, and compare its performance with manual methods.

Approach:
    Key Findings:
    • VisionMetric Suite achieved a recognition rate of 94.7%.
    • Average processing time was 2.8 ± 0.4 seconds per image.
    • User satisfaction among 10 physicians averaged 4.3/5.0.
    • Agreement between automated and manual measurements showed a mean ICC of 0.94 (95% CI: 0.91–0.97).
    • Repeatability of AI measurements showed a mean ICC of 0.96 (95% CI: 0.92–0.98).
    Interpretation:

    The preliminary validation suggests that VisionMetric Suite can serve as a rapid, objective periocular measurement tool with good agreement compared to manual methods.

    Limitations:
    • The study is a preliminary single-center validation.
    • Further validation in diverse clinical settings is needed.
    Conclusion:

    VisionMetric Suite represents an initial step toward automated assessment in ophthalmic plastic surgery.