Objective:

Develop a microscopy system that produces histology-like images and quantitative molecular maps from fresh, unprocessed tissue without sectioning or staining.

Approach:
  • Imaging method: A deep-UV microscope using LEDs measured tissue absorption at 255 and 280 nm and natural tryptophan fluorescence.
  • Image generation: Researchers used a mathematical transformation to create images resembling H&E-stained sections alongside maps of molecular composition.
  • Demonstration: The system was tested on fresh mouse brain and kidney tissue and human brain tumor specimens.
Key Findings:
  • The system resolved kidney structures including glomeruli and showed densely packed nuclei in human glioma samples.
  • Tumor regions had greater nuclear density and nucleic acid content than noncancerous brain tissue.
  • Tryptophan fluorescence efficiency was more than 50% lower in tumor tissue than in healthy human brain tissue.
  • Lateral resolution was approximately 0.5 µm; imaging a 10 × 10 mm area took about three minutes per wavelength.
Interpretation:

The study presents an early technical demonstration of slide-free imaging and quantitative molecular mapping in fresh tissue. The authors describe potential use for faster pathological assessment during surgery while preserving tissue for further testing.

Limitations:
  • Testing involved limited tissue types and only two human tumor specimens.
  • Clinical studies are needed to establish diagnostic accuracy, reproducibility, and performance across diseases.