ICU-led remote vital sign trends monitoring of selected non-ICU-patients: a first year experience - Summary - MDSpire
Coming Soon: Introducing MDSpire News. Learn more
Conexiant’s news site is now MDSpire News. Learn more

First-Year Experience With ICU-Led Remote Vital Sign Trend Monitoring in Selected Non-ICU Patients

  • By

  • Mark V. Koning

  • Roel V. Peelen

  • October 29, 2026

Share

Objective:

The article describes the implementation and first-year experience of wearable-based remote monitoring for selected non-ICU patients, with intensivists responsible for patient selection and ICU staff reviewing vital-sign trends.

Approach:
  • Patient selection: Intensivists selected cooperative patients who did not require ICU admission but were considered at risk of deterioration expected to develop over more than 2 hours, with ward care demands remaining manageable.
  • Monitoring: An ICU nurse reviewed wearable trends every 2 hours and visited each patient once per shift; an ICU physician reviewed trends every 4 hours and visited daily. Ward staff continued MEWS assessments every 4 hours.
  • Measures: Wearables continuously recorded oxygen saturation, 3-lead ECG, and respiratory rate; non-invasive blood pressure was measured hourly. Alarms were deactivated.
Key Findings:
  • From December 2024 to March 2026, 92 patients were monitored for 1,797 hours, with a 95.4% data capture rate; monitoring covered 19 wards and used no more than three wearables concurrently.
  • Severe infection (37 patients), pulmonary embolism (15), and COPD (9) were the most frequent indications. Median monitoring duration was 1 day.
  • Remote monitoring identified deterioration in 11 patients who subsequently required ICU admission. Seven additional ICU admissions were not detected by monitoring and occurred for other clinical or care-related reasons.
  • Four patients received ward treatment, including diuretics, anti-arrhythmics, or naloxone. The naloxone case involved unconsciousness without vital-sign abnormalities and was not detected by monitoring.
  • There were no in-hospital cardiac arrests, Rapid Response Team calls, emergency ICU admissions, or other ward-based interventions in the monitored cohort.
  • The authors report that vital-sign abnormalities were absent in 39% of deterioration cases, highlighting the role of patient visits alongside trend monitoring.
  • Initial equipment and installation costs were €52,332 for the central monitoring system and workstations, plus €22,020 for four reusable wearables; annual technical support cost €4,477. No additional staff were required.
Interpretation:

The authors describe ICU-led remote monitoring as a hospital-wide approach for selected patients, with bedside assessment remaining necessary because deterioration may occur without vital-sign abnormalities. They note potential advantages and disadvantages, including preserving ICU beds and avoiding handovers, but also the possibility of delayed ICU admission and dispersing near-critical patients across wards.

Limitations:
  • Single-hospital experience limits transferability.
  • The retrospective evaluation has inherent limitations, and indications and outcomes may change over time.
  • Adherence to scheduled reviews and patient visits was not measured.
  • Selection was at the individual intensivist’s discretion and may have been influenced by ICU-bed occupancy.
  • Only monitored patients were evaluated; patients not selected for monitoring were not assessed.
Conclusion:

The authors state that broader evaluation is needed to understand the efficiency and usefulness of this monitoring approach.