In Situ Fenestration With the Electrified Wire Technique—Exploring the Role of Application Duration on Fenestration Size: A Case Report - Summary - MDSpire

Investigation of Application Duration's Impact on Fenestration Size Using the Electrified Wire Technique in Situ Fenestration: A Case Study

  • By

  • Silvia Iodice

  • George Apostolidis

  • Daour Yousef al Sarhan

  • Jose I. Torrealba

  • Nils Zabel

  • Tilo Kölbel

  • Giuseppe Panuccio

  • July 15, 2026

Share

Objective:

To highlight the importance of energy application duration during electrified wire in situ fenestration (EWISF) and its potential impact on procedural outcomes.

Approach:
  • Patient Case: An 89-year-old man with a 54-mm pararenal abdominal aortic aneurysm underwent endovascular repair using EWISF for left renal artery stenosis.
  • Procedure Details: The procedure involved catheterization, stent placement, and the application of radiofrequency energy to achieve graft perforation.
  • Energy Application: A prolonged application of radiofrequency energy (approximately 2 seconds at 40 W) was crucial for successful catheter passage through the fenestration.
Key Findings:
  • Prolonged application of radiofrequency energy improved the success of catheter advancement through the graft perforation.
  • The procedure was completed successfully with no residual endoleak after necessary adjustments.
Interpretation:

The study indicates that the duration of energy application during EWISF can affect procedural outcomes.

Limitations:
  • The case study is based on a single patient, limiting generalizability.
  • Technical challenges in applying EWISF to visceral vessels were noted, including difficulties in catheter advancement.
Conclusion:

The findings indicate that careful management of energy application duration is critical in EWISF procedures.