Objective:

To evaluate the effects of a novel UHMWPE sleeve on anchor pullout strength after an initial anchor loosening event.

Approach:
    Key Findings:
    • Initial baseline pullout strength for intra-operative condition was 258.216 ± 16.955 N.
    • Initial baseline pullout strength for post-operative condition was 181.695 ± 18.257 N.
    • Cyclic loading reduced anchor pullout strength to approximately 70% of the intra-operative value.
    Interpretation:

    Limitations:
    • Study used synthetic bone, which may not fully replicate human bone properties.
    • Only specific anchor types and sizes were tested.