Investigators found comparable measurement reliability across 2 analysis tools, but an 8-percentage-point difference in their results raises concerns about interchangeability.
In 93 patients with open-angle glaucoma, OCTAVA estimated macular vessel density 8.42 percentage points higher than the Erlangen-Angio tool.
2
Both tools showed good, comparable repeatability for average vessel density, with intraclass correlation coefficients of 0.85 for OCTAVA and 0.80 for Erlangen-Angio.
3
The inferior macular sector had the greatest measurement variability, with repeatability coefficients of 9.53 percentage points for OCTAVA and 7.65 for Erlangen-Angio.
4
Higher average macular vessel density was associated with better central visual field sensitivity for both tools after adjustment for image quality.
5
The authors stated that absolute values should not be compared across platforms and that the tools cannot be used interchangeably to monitor glaucoma progression.