    The study evaluated analytical agreement between Atellica HEMA 580 and Sysmex XN-1000 analyzers using 100 residual K₂EDTA whole-blood samples.

  • 2

    Strong analytical agreement was found for RBC, HGB, WBC, and PLT-I, with correlation coefficients exceeding 0.98 for these parameters.

  • 3

    MCV and RDW-CV showed analyzer-dependent differences, with MCV exceeding allowable bias limits and RDW-CV demonstrating lower correlation.

  • 4

    Bland–Altman analysis confirmed small mean bias for RBC and HGB, while wider limits of agreement were observed for WBC and PLT-I.

  • 5

    The Atellica HEMA 580 can be applied in clinical practice, but caution is advised when interpreting MCV and RDW-CV results across platforms.