    The study evaluated the accuracy of the Watch-PAT device for diagnosing sleep apnea in patients with chronic insomnia.

  • 2

    44 participants were analyzed, showing a sensitivity of 77.3% and specificity of 100% for identifying moderate-to-severe obstructive sleep apnea.

  • 3

    Diagnostic agreement was achieved in 82% of subjects when comparing Watch-PAT results with polysomnography.

  • 4

    The Watch-PAT device overestimated total sleep time and REM sleep duration by 19.9 minutes and 37.2 minutes, respectively.

  • 5

    The study highlights the need for cautious interpretation of sleep duration and staging parameters from the Watch-PAT device.