Quantification of tension during cruroplasty of robotic-assisted hiatal hernia repairs using force feedback instruments on the da Vinci surgical platform (dV5) - Takeaways - MDSpire
Coming Soon: Introducing MDSpire News. Learn more
Conexiant’s news site is now MDSpire News. Learn more

Measuring Tension During Robotic Hiatal Hernia Repair Crural Closure with Force-Feedback Instruments on the da Vinci dV5 Platform

  • By

  • Khang Ngo

  • Marzieh Ershad Langroodi

  • Prakash Gatta

  • October 12, 2026

Share

  • 1

    The study assessed whether da Vinci 5 force-feedback instruments could quantify crural closure tension and crural stiffness during robotic hiatal hernia repair.

  • 2

    Crural closure tension includes axial force related to esophageal shortening and radial force across the hiatus, while crural tissue compliance is currently assessed subjectively.

  • 3

    The retrospective, single-institution cohort included 76 patients undergoing primary or revisional robotic repair between January 2024 and December 2025.

  • 4

    Force measurements were retrieved from system logs and were not available to surgeons during operations; 32 of 108 cases were excluded for missing or unreliable data.

  • 5

    The exploratory Crural Stiffness Index was defined as peak force per centimeter of crural displacement, with clinical validation and decision support identified as future research stages.

Original Source(s)