    Thulium fiber lasers (TFL) and holmium: YAG lasers (Ho: YAG) are widely used in urology for various procedures.

  • 2

    The MAUDE database revealed that device malfunctions were more common with Ho: YAG lasers compared to TFLs.

  • 3

    Patient adverse events were significantly higher with Ho: YAG lasers, with 35.3% compared to 13.5% for TFLs.

  • 4

    After the 2021 FDA recall, TFL-related events showed a significant decrease in patient harm and fiber breakages.

  • 5

    Most adverse events for both laser types were classified as minor, but TFLs had a higher proportion of Level I events.