    The study compares the biomechanical stability of clavicular hook plates and adjustable strap plates for Neer Type IIB distal clavicle fractures.

  • 2

    Finite element analysis showed both fixation methods provided effective stability, with no fracture risk for the internal fixation materials.

  • 3

    Model A (hook plate) had a smaller maximum deformation (0.54 mm) compared to Model B (0.79 mm), but no significant difference in mean deformation.

  • 4

    Model A exhibited higher mean equivalent stress (3.42 MPa) than Model B (2.87 MPa), indicating different stress distribution characteristics.

  • 5

    The adjustable strap plate may offer advantages over the hook plate, including reduced complication risks and simultaneous management of associated injuries.