Taking the Pressure Off Tonometry

  • September 17, 2026

  • 3 min

Share

  • 1

    A handheld, automated applanation tonometer has shown accurate intraocular pressure measurements in human donor eyes.

  • 2

    The device automates the subjective component of traditional Goldmann applanation tonometry, improving measurement consistency.

  • 3

    Testing on eight human cadaveric eyes demonstrated a mean difference of −1.2 mm Hg from set intraocular pressure values.

  • 4

    Performance was notably better in the 10–30 mm Hg range, with a mean absolute difference of 1.4±1.1 mm Hg.

  • 5

    The study highlights the need for further clinical evaluations to assess usability and performance across diverse corneas.